SPECIFICATIONS
Measurement Units:
µ Mole ZPP/Mole Heme (S.I. units)
(No hematocrit correction required).
Alternate Measurement Units:
1. µg ZPP/gram Hb
2. EQ FEP µg 100 ml whole blood, Hct 35 (CDC)
3. µg ZPP/100 ml whole blood, Hct 42 (OSHA)
Display:
3 1/2 digit L.E.D.
Non-Linearity:
<1%
Reproducibility:
±1% full scale or 2 least significant digits,
whichever is greater.
Accuracy:
.98 (100 sample correlation with Free Erythrocyte Protoporphyrin). Hematofluorometer controls, stable for 3 months at -20°C, (low, medium, high).
Power Requirements:
117 VAC ± 10%, 60 Hz, 220 VAC ± 10%, 50Hz.
Dimensions:
9” W x 12”H x 15”D (22.9W x 30.4H x 38.1D cm.)
Weight:
20 lbs. (9 kg.)
|